DC Power Supply Models AX501, AX502 and AX503

 

3

 

CHAPTER 1

INTRODUCTION

  WARNING 

These safety warnings are provided to ensure the safety of personnel 

and proper operation of the instrument.

• Read this user manual completely and follow all the safety information

before attempting to use or service this instrument.

• Use  caution  on  any  circuit:  Potentially  high  voltages  and  currents

may be present and may pose a shock hazard.

• Only use leads with the proper rating. Always inspect the instrument

and lead prior to use. Replace any defective parts immediately.

• Safety is the responsibility of the operator!
• Any defective lead insulation, inside or outside of the instrument, or

disconnection of the protective earth/ground terminal may make the

instrument dangerous.

• The  plug  should  only  be  inserted  into  a  socket  equipped  with  a

grounding  contact.  The  safety  connection  must  not  be  broken  by

use of an extension lead without protective insulation.

• Before connecting the power supply, make sure the distribution net

-

work is 115V

AC

 ±10%.

1.1 

International Electrical Symbols

This symbol signifies that the instrument is protected by double or 

reinforced insulation. Use only specified replacement parts when 

servicing the instrument.
This  symbol  on  the  instrument  indicates  a  WARNING  and  that 

the operator must refer to the user manual for instructions before 

operating  the  instrument.  In  this  manual,  the  symbol  preceding 

instructions  indicates  that  if  the  instructions  are  not  followed, 

bodily injury, installation/sample and product damage may result.
Risk of electric shock. The voltage at the parts marked with this 

symbol may be dangerous.
Earth/Ground

www.

ShopAEMC

.com

Shop for AEMC products online at:

1.888.610.7664

Summary of Contents for AX501

Page 1: ...100V CAT I V A 30V LIMIT 0 COARSE FINE 2 5A 0 IEC 61010 1 POWER ON OFF AX 503 8 8 8 8 8 8 8 8 8 8 8 8 MASTER TRACKING SLAVE 100V CAT I 100V CAT I 100V CAT I A V A V V 2 5A 30V 3V 5V 2 7V 5 5V LIMIT 5A...

Page 2: ...nt to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibration...

Page 3: ...Voltage and Current Adjustments 9 2 3 3 Current LIMIT LED 9 2 3 4 TRACKING Switch Models AX502 and AX503 10 2 3 5 Setting TRACKING 10 2 3 6 Digital LED displays 11 2 3 7 Output Terminals 11 2 3 8 Gro...

Page 4: ...Power Supply AX503 Only 18 5 MAINTENANCE 19 5 1 Warning 19 5 2 Fuse Replacement 19 5 3 Cleaning 19 5 4 Storage 19 Repair and Calibration 20 Technical and Sales Assistance 20 Limited Warranty 21 Warra...

Page 5: ...rous The plug should only be inserted into a socket equipped with a grounding contact The safety connection must not be broken by use of an extension lead without protective insulation Before connecti...

Page 6: ...he contents are consistent with the packing list Notify your distributor of any missing items If the equip ment appears to be damaged file a claim immediately with the carrier and notify your distribu...

Page 7: ...housing make them top level laboratory instruments The AX501 offers Single output Simultaneous display of the output voltage and current Voltage adjustable from 0 to 30V for a variable current of 0 to...

Page 8: ...ge display green 2 Current display red 3 Current limit LED 4 Power On Off button 5 Fine voltage adjustment potentiometer 6 Output terminal 7 Output terminal 8 Coarse voltage adjustment potentiometer 9...

Page 9: ...limit LED 3 Voltage display green 4 Tracking function LED 5 Tracking On Off switch 6 Current limit adjustment potentiometer 7 Power On Off button 8 Fine voltage adjustment potentiometer 9 Output termi...

Page 10: ...ay green 4 Tracking function LED 5 Tracking On Off switch 6 Current limit adjustment potentiometer 7 Power On Off button 8 Fine voltage adjustment potentiometer 9 Output terminals 10 Output terminals...

Page 11: ...left decreases output The current value displayed is the value of the output current on the red display 2 3 3 Current LIMIT LED The AX Series Power Supplies provides the ability to limit the current...

Page 12: ...output You need to set this before turning on the Tracking To set up the Tracking first set the MASTER voltage output then set the SLAVE output To set this simply adjust the SLAVE output voltage NOTE...

Page 13: ...rs with a resolution of 0 1V and 10mA 2 3 7 Output Terminals The female two pin and safety terminals 4mm are used to connect the outputs to the points of the circuit to be powered by means of safety l...

Page 14: ...to 5 5 V Adjustment coarse and fine by potentiometers by potentiometer Display green LEDs none Resolution 0 1 V Accuracy 0 5 of reading 1 digit Stability if variation mains 10 of the 0 to 2 5 A load...

Page 15: ...ction against overheating yes 3 2 General Specifications Power Supply 115VAC 10 Consult the factory for changing to 220VAC Display Dual display VOLTAGE in Green and CURRENT in Red 7 segments 0 56 14 2...

Page 16: ...uts Overvoltage Cat 100V CAT I Max Output Voltage 30 5VDC in normal mode 61 0VDC in series mode Power supply inputs Overvoltage CAT 300V CAT II Supply Current 115V 10 50 60Hz Electromagnetic Compatibi...

Page 17: ...easuring circuits make sure that you are not charged with static electricity which could irreparably damage the instrument s internal components WARNING When the instrument is open some of the interna...

Page 18: ...sing values The voltage set point is then dis played Adjust the value of the current limit by turning the current control to the right increasing values The current value displayed is the value of the...

Page 19: ...be powered V max 60V Current max 2 5A ON Sample Figure 5 To use 2 outputs in series with TRACKING symmetrical 30V sources at 2 5A max see Fig 6 Procedure for serial number 105500MAV Insequence powersu...

Page 20: ...utside its operating limits voltage or current by the MASTER source the corresponding LIMIT LED lights up 4 2 4 Use of the 2 7V to 5 5V 5A Power Supply AX503 Only If necessary connect the earth ground...

Page 21: ...power supply to change the fuse you must disconnect the leads and power cord Only use a fuse of the following type 10A 250V 5 x 20mm 5 3 Cleaning Disconnect the instrument then clean it with a cloth l...

Page 22: ...e the CSA on the outside of the shipping container If the instrument is returned for calibration we need to know if you want a standard calibration or a calibration traceable to N I S T Includes calib...

Page 23: ...at its option repair or replace the faulty material Warranty Repairs What you must do to return an Instrument for Warranty Repair First request a Customer Service Authorization Number CSA by phone or...

Page 24: ...03 19 99 MAN 100252 v13 www ShopAEMC com Shop for AEMC products online at 1 888 610 7664...

Reviews: