5
FUNCTION SELECTOR SWITCH
OFF
Off position
V~
AC voltages up to 600V in rms value
V
...
DC voltages up to 600V
Hz
Frequencies of voltages up to 400kHz
Ω
Ω
Ω
Ω
Ω
Continuity sound test for a resistance R
≤
40
Ω
, and
measurements of resistances up to 40M
Ω
Note:
The continuity test
is selected automatically
when the instrument is switched on.
Diode test. Measurement of the voltage of the junction in
forward direction
Capacitances up to 40µF
mA
Current ranges up to 400mA
AC/DC
(via the + terminal)
A
10A
AC/DC
current ranges (via the 10A terminal)
TERMINALS
Ø 4mm safety terminals
COM:
common, black
+:
red
10A:
red lead for the 10A
AC/DC
ranges
3 - DC AND AC VOLTAGE
Connect the leads to the multimeter and connect it in parallel to the
circuit to be tested.
Place the selector switch on the voltmeter function V
AC/DC
Automatic selection of the range: read the value measured.
Freeze the value by pressing the HOLD button.
If necessary, press the following buttons:
RANGE to select the ranges manually
MIN-MAX to read the min or max values
to light the display
V
AC/DC
400mV
(1)
4V
40V
400V
600V
(2)
Digital
0.1mV
1mV
10mV
100mV
1V
resolution
Bargraph
10mV
100mV
1V
10V
100V
resolution
Impedance
100M
Ω
10M
Ω
9.1M
Ω
Accuracy
± 0.25% of Reading ±1ct
on DC
Accuracy
± 1.5% R
± 1% of Reading ±3cts
on AC
±5cts
Permitted
600Vrms and 1000V peak
overload
(1) Accessible by the RANGE button. Min measurement : 20mV
(2) Crest factor FC
≤
3