background image

SPECIFICATIONS

Analog Inputs

8 analog inputs:

Thermocouple J, K, T, E, N, R, S and B (IEC-584)

Pt100, Pt1000 (IEC-751), 0 to 20mA, 4 to 20mA,
0 to 20mV, 0 to 50mV, 0 to 60mV, - 20 to 20mV,
0 to 5V and 0 to 10V

Analog Accuracy

Thermocouple J, K, T, E and N: 0.2% of range
± 1°C

Thermocouple R, S and B: 0.2% of range ± 3°C

Pt100, Pt1000, 0 to 20mA, 4 to 20mA, 0 to 20mV, 0
to 50mV, 0 to 60mV, - 20 to 20mV, 0 to 5V and 0
to 10V: 0.15% of maximum range

Digital Inputs /

Outputs

8 digital channels that can be individually configured
as inputs or outputs

Relay Outputs

2 Relays SPST-NO/NC, 3A / 250V

AC

RS485 Interface

Can be used as a Modbus RTU slave or master (can
read up to 64 slaves)

Ethernet Interface

10/100 Mbps

Supported

Protocols and

Services

TCP/IP, DHCP, HTTP, SMTP, SNMP, Modbus RTU,
Modbus TCP, FTP server and client

Internal Memory

2 MB

SD Card

Up to 16 GB

Operating

Environment

32 to 122

°

F (0 to 50°C). Relative Humidity

(Maximum): 80% up to 86

°

F (30ºC). For

temperatures above 86

°

F (30ºC), decrease 3% per

°

F/ºC. Installation category II. Pollution degree 2.

Altitude < 2000 m

Enclosure

ABS + PC

Dimensions

6.5 x 4.6 x 2.8" (165 x 117 x 70mm); Weight: 400 g.
(no HMI)

Power Supply

100 to 240V

AC/DC

(50/60Hz)

24V

AC/DC

(50/60Hz)

Protection

IP20

Configuration

Software Operation

Environment

Software for Windows

®

XP, Vista and 7

Menus in Portuguese, Spanish or English

Table 01

DATA LOGGER

DL-1080 / DL-1081

Q U I C K

S T A R T

G U I D E

QG 5000967 – 99-MAN 100384