Megohmmeter Model 6522/6524
13
Model 6524:
The alarm is available in insulation, resistance, and continuity
measurement modes. Pressing the
button activates the
alarm. The symbol is displayed, along with the threshold
value.
ALARM
<
To turn OFF the alarm buzzer while it is sounding, press the
HOLD
button. To
deactivate an active alarm function, press the button.
2.1.2.2 Setting an Alarm Threshold
While
is displayed indicating the alarm function is active, you can change
the alarm threshold by pressing the
button (except during insulation
measurements). For each testing mode, there are three pre-defined threshold
values:
■ Continuity: < 2Ω, <1Ω and <0.5Ω
■ Resistance: >50kΩ, >100kΩ and >200kΩ
■
Insulation:
50V : <50kΩ, <100kΩ and <20kΩ
100V : <100kΩ, <200kΩ and <400kΩ
250V : <250kΩ, <500kΩ and <1MΩ
500V : <500kΩ, <1MΩ and <2MΩ
1000V : <1MΩ, <2MΩ and <4MΩ