Megohmmeter Model 6522/6524
25
M
mA
R
M
mA
The instrument displays the
resistance and the current used in
the test.
To obtain a continuity value per
standard IEC 61557:
1.
Take a measurement at
200 mA and note its value,
R1.
2.
Reverse the leads and
note the value R2.
3.
Calculate the mean:
2.2.4 Resistance Measurement (Model 6524)
Resistance measurements up to 1000k
Ω
are made with a low current.
1.
Set the rotary switch to
k
Ω
.
100V
V
k
Ω
MR
OFF
Ω
250V
500V
1000V
50V
M
Ω