Megohmmeter Model 6522/6524
29
3. SPECIFICATIONS
3.1 General Reference Conditions
Quantity of Influence
Reference Values
Temperature
23° ± 3°C (73.4° ± 5.4°F)
Relative humidity
45 to 55% RH
Frequency
DC and 45 to 65Hz
Supply voltage
8 ± 0.2V
battery life indication 58% ± 8%
Electric field
0V/m
Magnetic field
< 40A/m
■
The intrinsic uncertainty
is the error specified for the reference
conditions.
■
The operating uncertainty
includes the intrinsic uncertainty plus
variations of the quantities of influence (position, supply voltage,
temperature, etc.) as defined in standard IEC 61557.
In this section, uncertainties are typically expressed as % of the reading (R) plus
number of display counts (ct).
3.2 Electrical Specifications
3.2.1 Voltage Measurement
Specific reference conditions: Peak factor = 1.414 in AC, sinusoidal signal.
Measurement Range
0.3 to 399.9V
400 to 700V
Resolution
0.1V (AC and DC)
1V (AC and DC)
Intrinsic uncertainty
± (3% R + 2 ct)
Input impedance
400kΩ
Frequency ranges
DC and 15.3 at 800Hz