Megohmmeter Model 6522/6524
31
Test Voltage
Measurement
Range
0.0 to 399.9V
400 to 1250V
Resolution
0.1V
1V
Accuracy
± (3% R + 3 ct)
Typical Discharge Time after Test
To go from V
T
to 25V, the discharge time is < 2s/µF.
Test Current
Maximum test current: 2mA
Measurement
Range
0.01 to
39.99µA
40.0 to
399.9µA
0.400 to
2.000mA
Resolution
10nA
100nA
1µA
Accuracy
± (10% R + 3 ct)
Typical test voltage vs load curve
The voltage as a function of the measured resistance is illustrated below:
The range of operation per IEC 61557 is from 100kΩ to 2GΩ (see § 3.3.4
).