3.2.4 Continuity Measurement
Specific reference condition: Inductance in series with the resistance = zero.
Measurement Range (without
compensation of the leads)
0.00* to 10.00Ω
0.0 * to 100.0Ω
Resolution
10m
Ω
100m
Ω
Accuracy
± (2% + 2 ct)
Test Current
200mA
20mA
Open Voltage
≥
6V
*In the case of incorrect compensation of the leads, the instrument allows display
of negative values, down to -0.05Ω at 200 A and -0.5Ω at 20mA.
Test current
200mA range: 200mA (0mA + 20mA)
20mA range: 20mA ± 5mA
Measurement Range
0 to 250mA
Resolution
1mA
Accuracy
± (2% + 2 ct)
Compensation of the leads
: 0 to 9.99Ω.
3.2.5 Resistance Measurement (Model 6524)
Measurement Range
0 to
3999Ω
4.00 to
39.99kΩ
40.0 to
399.9kΩ
400 to
1000kΩ
Resolution
1
Ω
10
Ω
100
Ω
1k
Ω
Accuracy
± (3% + 2 ct)
Open voltage
approximately 4.5V
3.2.6 Timer
Measurement Range
0:00 to 39:59
Resolution
1s
Accuracy
± 1s