HWS.1 Handy Wire Scanner - 240HW01V00
B2798_E 1.1, 2016-04-08
23
•
Power Management
It Is used to select the downtime after which the instrument is switched off automatically
PLEASE REFER TO CHAPTER 2.10 AUTOMATIC SWITCH OFF
Selecting with the
Enter key
(
F1
) you access the item
Power Off
and selecting again the entry with the
Enter key
(
F1
) you access to the choice of the no-activity time (downtime) that must pass before the instrument will
automatically switch off. You can choose between the following values:
OFF (NO automatic power off), 30 sec, 1 min, 2 min, 3 min, 5 min, 10 min, 15 min
Select the desired value using the
arrow keys
(
F2 and F3
) and the
Enter key
(
F1
).
•
Language
It is used to select language for the menus of the instrument.
Selecting the item with the
Enter key
(
F1
) you access the menu to choose from the following options:
English, Italiano
Select the desired value using the
arrow keys
(
F2
and
F3
) and the
Enter key
(
F1
).
•
Calibration
It is used to correct the calibration of the instrument, setting a fixed offset value (positive or negative) to be added
to the measurement made by the instrument.
Selecting with the Enter key (F1) you access the items Offset, Diameter and Factory. At the side of voice Offset, the
value of the offset in use is shown.
Offset
: If you select the entry with the Enter key (F1) you access the window for adjusting directly the offset
value.
Diameter
: If you select the entry with the Enter key (F1) you to access the window for setting the offset value by
correcting the diameter value measured using a calibration master
Factory
: If you select the entry with the Enter key (F1) you access the menu to restore the factory calibration
(Offset = 0)
For all information relating to Calibration:
PLEASE REFER TO CHAPTER 3.7 OFFSET CALIBRATION
•
Reserved Aeroel
It is used to access a service menu reserved to the manufacturer