MAINTENANCE MANUAL
IFR 4000
2-2-3
Page 62
Aug 1/04
STEP PROCEDURE
LOC
8. Press the LOC Soft Key to display the LOC AMPL LEVEL Sequence Screen.
9. Connect Measuring Receiver Sensor Head to the ANT Connector.
10. Follow the on-screen prompts and recall stored calibration levels on the
Measuring Receiver (para 2-2-3F[1] Step 2). For the LOC AMPL LEVEL
Sequence, the AM modulation setting error must be <0.5% and the amplitude
setting error must be <0.1 dB.
11. Reference the following:
z
Step 1 of LOC AMPL LEVEL Sequence
Turn modulation OFF to set and check RF level. Toggling the UUT
Modulation field between OFF and ON and the Measuring Receiver between
AM and TUNED RF LEVEL several times may be required to meet the
requirements of this step.
z
Step 2 of LOC AMPL LEVEL Sequence
Once the VALUE: field has been set for minimum distortion, use the PREV
Soft Key to return to Step 1. Steps 1 and 2 are interactive; toggling between
the two is required to meet the requirements of the steps.
12. When the LOC AMPL LEVEL Sequence is complete, press the SAVE & RETURN
Soft Key to return to the AMPL LEVEL Cal Screen.
MKR
13. Press the MKR Soft Key to display the MKR AMPL LEVEL Sequence Screen.
14. Connect 10 dB Attenuator between Measuring Receiver Sensor Head and ANT
Connector.
15. Follow the on-screen prompts and recall stored calibration levels on the
Measuring Receiver (para 2-2-3F[1] Step 2). For the LOC AMPL LEVEL
Sequence, the AM modulation setting error must be <1% and the amplitude
setting error must be <0.1 dB (taking into account the 10 dB Attenuator).
16. Reference the following:
z
Step 1 to 2 of MKR AMPL LEVEL Sequence
Turn modulation OFF to set and check RF level. Toggling the UUT
Modulation field between OFF and ON and the Measuring Receiver between
AM and TUNED RF LEVEL several times may be required to meet the
requirements of this step.
z
Step 3 of MKR AMPL LEVEL Sequence
Select 15 kHz Low-Pass Filter on the Measuring Receiver.
17. When the MKR AMPL LEVEL Sequence is complete, press the SAVE & RETURN
Soft Key to return to the AMPL LEVEL Cal Screen.
18. Press the RETURN Soft Key to return to the Calibration Screen.
Summary of Contents for IFR 4000
Page 1: ...NAV COMM Test Set Maintenance Manual 1002 5600 4P0 IFR 4000...
Page 3: ...MAINTENANCE MANUAL IFR 4000 FOR QUALIFIED SERVICE PERSONNEL ONLY...
Page 4: ...MAINTENANCE MANUAL IFR 4000 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 6: ...MAINTENANCE MANUAL IFR 4000 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 12: ...MAINTENANCE MANUAL IFR 4000 INTRODUCTION Page 2 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 32: ...MAINTENANCE MANUAL IFR 4000 2 2 1 Page 14 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 34: ...MAINTENANCE MANUAL IFR 4000 2 2 1 Page 16 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 42: ...MAINTENANCE MANUAL IFR 4000 2 2 2 Page 8 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 108: ...MAINTENANCE MANUAL IFR 4000 2 2 4 Page 2 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 160: ...MAINTENANCE MANUAL IFR 4000 2 2 4 Page 54 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 166: ...MAINTENANCE MANUAL IFR 4000 2 3 1 Page 6 Aug 1 04 STEP PROCEDURE 4 Remove the Fuse...
Page 186: ...MAINTENANCE MANUAL IFR 4000 APPENDIX B Page 2 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 188: ...MAINTENANCE MANUAL IFR 4000 APPENDIX C Page 2 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 200: ...MAINTENANCE MANUAL IFR 4000 APPENDIX D Page 12 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...
Page 206: ...MAINTENANCE MANUAL IFR 4000 APPENDIX E Page 6 Aug 1 04 THIS PAGE INTENTIONALLY LEFT BLANK...