Aeroneb
®
Lab Nebulizer System Instruction Manual
5
Warnings
Read and study all instructions before using the Aeroneb
®
Lab. Only trained personnel should operate
the device.
For pre-clinical use only. It is not intended for use with humans.
Do not use in the presence of a flammable anesthetic mixture combined with air or with oxygen or
nitrous oxide.
To avoid the risk of fire, do not use in the presence of flammable substances.
Assemble and perform a functional test according to the instructions in this manual before use.
If this equipment is not used in a manner specified in this manual, the protection provided by the
equipment may be impaired.
The ability to effectively aerosolize a novel formulation or chemical entity while maintaining the
integrity of this entity can only be determined through direct testing, and is the responsibility of the
device user.
Cautions
To avoid mechanical or electrical damage, do not drop the nebulizer unit or the control module.
Disconnect nebulizer unit from control module before cleaning.
Do not immerse in water or autoclave the control module, cables or AC/DC adapter.
Use only with components specified by Aerogen.
Inspect all parts before use, and do not use if any parts are missing, cracked or damaged. In case of
missing parts, malfunction or damage, contact your Aerogen representative.
Do not use or store outside of specified environmental conditions.
Electromagnetic Susceptibility
This device meets the requirements for Electromagnetic Compatibility as defined by European
Directive 89/336/EEC relating to Electromagnetic Compatibility. This includes immunity to radio
frequency electric fields and electrostatic discharge, in addition to the other applicable requirements of
the standard. Compliance with EMC standards does not mean a device has total immunity; certain
devices (cellular phones, pagers, etc.) can interrupt operation if they are used near the equipment.