16
Aeroneb
®
Solo System Instruction Manual
5.
To operate on AC power (the primary mode of
operation), insert the AC/DC adapter cable into the
control module and plug the adapter into an AC power
source.
6.
Aeroneb
®
Pro-X can be battery-operated for portable
applications. The rechargeable battery can power
the System for up to 45 minutes when fully charged.
In the case of AC power failure the control module will
automatically switch to battery operation.
Note:
Allow a minimum of eight hours for the internal battery
to fully recharge.
Note:
To ensure uninterrupted operation of Aeroneb
®
Solo,
secure both the AC/DC adapter cable and the control module
cable so they cannot become disconnected during treatment.
If clips are available on patient circuits, run the cables through
the eyes of the clips. If clips are not available, ensure that all
cables are routed safely.
Note:
The continuous mode can only be operated from AC
power supply. The AC/DC adapter is the means of isolating
the Aeroneb
®
Solo system from the main power supply.