Functional test
Perform a functional test of the Aeroneb
®
Solo System prior to
first use or at any time to verify proper operation. Follow these
steps:
1.
Visually inspect each part of the System for cracks or
damage and replace if any defects are visible.
2.
Pour 1-6 mL of normal saline (0.9%) into the
nebulizer unit.
3.
Connect the nebulizer unit to the control module using
the control module cable. Connect the control module
to the AC/DC adapter and plug the AC/DC adapter into
an AC power source.
4.
Press and release the blue on/off power button and
verify that the green 30
Min.
indicator LED illuminates
and that aerosol is visible.
5.
Press the on/off power button again to turn the
System off. Press and hold the button for at least three
seconds. Verify that the green
Continuous
indicator
LED illuminates and that aerosol is visible.
6.
Disconnect the control module from the AC/DC adapter
and verify that nebulization continues and that the
battery status indicator turns off.
7.
Turn the System off and verify that the 30
Min.
and
Continuous
indicators are off.
Summary of Contents for Solo System
Page 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Page 2: ......
Page 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Page 4: ...This page has been intentionally left blank...
Page 46: ...This page has been intentionally left blank...
Page 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Page 48: ...This page has been intentionally left blank...
Page 55: ......