Aeroneb
®
Solo System Instruction Manual
Recommended separation distances between portable and
mobile RF communication equipment and the Aeroneb Solo
The Aeroneb Solo nebulizer system is intended for use in the
electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Aeroneb Solo
can help prevent electromagnetic interference by maintaining a
minimum distance between port and mobile RF communications
equipment (transmitters) and the Aeroneb Solo nebulizer system as
recommended below, according to the maximum output power of
the communications equipment.
Rated
maximum
output power
of transmitter
W
Separation distance according to frequency
of transmitter
m
150 kHz to
80 MHz
d = [1.17] √P
80 MHz to
800 MHz
d = [1.17] √P
800 MHz to
2.5 GHz
d = [2.33] √P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.75
1
1.17
1.17
2.33
10
3.70
3.70
7.36
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Summary of Contents for Solo System
Page 1: ...Rx Only US 8 30 mins Symbols for Page 8 and 9 Instruction Manual...
Page 2: ......
Page 3: ...Part Number AG AS3050 EN Rev J 2011 Aerogen Ltd Instruction Manual Aeroneb Solo System...
Page 4: ...This page has been intentionally left blank...
Page 46: ...This page has been intentionally left blank...
Page 47: ...Aeroneb Solo System Instruction Manual 41 Appendix 1 EMC tables...
Page 48: ...This page has been intentionally left blank...
Page 55: ......