Overview
ABL1000WB Hardware Manual
1.1. Environmental Specifications
W A R N I N G : Do not expose this product to environments or conditions outside of the listed
specifications. Exceeding environmental or operating specifications can cause damage to the
equipment.
Table 1-2:
Environmental Specifications
Ambient
Temperature
Operating: 16° to 25° C (61° to 77° F)
The optimal operating temperature is 20° C ±2° C (68° F ±4° F). If at any time the
operating temperature deviates from 20° C degradation in performance could occur.
Storage: 0° to 40° C (32° to 104° F) in original shipping packaging
Humidity
Operating: 40 percent to 60 percent RH
The optimal operating humidity is 50 percent RH.
Storage: 30 percent to 60 percent RH, non-condensing in original packaging
Altitude
Operating: 0 m to 2,000 m (0 ft to 6,562 ft) above sea level
Contact Aerotech if your specific application involves use above 2,000 m or below sea
level.
Vibration
Use the system in a low vibration environment. Excessive floor or acoustical vibration
can affect system performance. Contact Aerotech for information regarding your
specific application.
Protection
Rating
ABL1000WB stages are not suited for dusty or wet environments. This equates to an
ingress protection rating of IP00.
Use
Indoor use only
1.2. Accuracy and Temperature Effects
Due to the small clearances in the air bearing design, extreme temperature environments could cause a
decrease in performance or permanent damage to the stage. Standard Aerotech air-bearing stages are
designed for and built in a 20°C (68°F) environment. The environmental temperature must be controlled to
within 0.25°C per 24 hours to ensure that the ABL1000WB specifications are repeatable over an extended
period of time. The severity of temperature effects on all specifications depends on many different
environmental conditions, including how the ABL1000WB is mounted. Contact the factory for more details.
12
Chapter 1
www.aerotech.com