ATTACH THE SCAN HEAD TO THE MOUNTING FLANGE
A general procedure for attaching the AGV-HP(O) scan head to a mounting flange is provided below.
Note that the mounting flange must include appropriate features to match the threaded holes and
the dowel holes and slots present on the AGV-HP(O) scan head.
1. Remove the plug from the beam entrance aperture of the scan head.
IMPORTANT
: Retain the plug for future use. Reinstall the plug if you need to ship the
scan head back to the factory for service.
2. Prepare the mounting flange and the mating surface of the scan head by stoning with precision
flatstones to remove any burrs or high spots.
3. Clean the mounting flange and the mating surface of the scan head with the appropriate cleaners
(isopropyl alcohol for the scan head).
4. Line up the dowel hole and slot features of AGV-HP(O) scan head with the dowel pins protruding
from the mounting flange.
5. Install the scan head onto the pins.
6. Insert the appropriate hardware into the holes of the mounting flange. The screw length must be
long enough to provide at least 1.0x diameter thread engagement.
7. Using an appropriate wrench, tighten the hardware. Alternate tightening of the screws so that the
AGV-HP(O) scan head comes together evenly on the mounting flange.
2.3. Securing the Scan Head to the Mounting Surface AGV-HP(O) Hardware Manual
42
www.aerotech.com