3.2. Motor and Feedback Wiring
DANGER
: To decrease the risk of electrical shock, injury, death, and damage to the
equipment, obey the precautions that follow.
l
Make sure that all components are grounded correctly and that they obey the local
electrical safety requirements.
l
It is the responsibilty of the system integrator or qualified installer to determine and
meet all safety and compliance requirements when they integrate the AGV-HP(O)
into a completed system.
l
Restrict access to the galvo when it is connected to a power source.
AGV-HP(O) scan heads are wired and assembled at the factory. Contact your Aerotech representative
if you have any questions on system configuration. For cable drawings, refer to the documentation
that shipped with your system.
Table 3-3:
Aerotech Motor and Feedback Cable Part Numbers
Aerotech Controller
Nmark GCL or Automation1 GL4
Cable Application
Standard
Hi-Flex
Interface Cable Part Number
C23680-xxx
(1,2)
C23690-xxx
(1,2)
(1) -xxx = Cable Length in Decimeters (1 Decimeter = 3.937 inches)
(2) Contact the factory for standard and custom length cables
Table 3-4:
Wire Gauge and Insulation Specifications
Current
Wire Gauge: mm
2
(AWG)
Minimum Recommended
Insulation Voltage (V)
AGV10HP(O)
7.1
1.3 (16)
60
AGV14HP(O)
7.1
1.3 (16)
60
AGV20HP(O)
7.1
1.3 (16)
60
AGV30HP(O)
5.6
1.3 (16)
60
3.2. Motor and Feedback Wiring
AGV-HP(O) Hardware Manual
50
www.aerotech.com