2.6. Air Requirements
WARNING
: To prevent damage to the AGV3D, do not attach a water hose to an air fitting
or an air hose to a water fitting.
DANGER
: High pressure air can cause severe injury.
l
Lock out the source and bleed off the pressure before you do service to the equip-
ment.
l
Wear eye protection.
If the -AC (air-cooling) option was purchased, use this option to cool the turning mirrors and dynamic
focusing module.
A gas supply pressure of 517 kPa to 586 kPa (75 psi to 85 psi) is required. Gas must be delivered via a
polyurethane air hose with an outer diameter of 4 mm. Unless otherwise specified, a gas flow rate of 20
SLPM (standard liters per minute) at 550 kPa (80 psi) should be observed.
l
If nitrogen is used, it must be 99.99% pure and filtered to 0.25 microns. [Recommended]
l
If compressed air is used, it must be filtered to 0.25 microns, dry to 0º F dew point, and oil free.
The filtration requirement is necessary to prevent particles from damaging the optical surfaces of the
turning mirrors.
Figure 2-7:
Air-Cooling Locations
2.6. Air Requirements
AGV3D Hardware Manual
40
www.aerotech.com