1.1. Environmental Specifications
WARNING
: Use this product only in environments and operating conditions that are
approved in this manual.
Table 1-2:
Environmental Specifications
Ambient
Temperature
Operating: 10° to 35° C (50° to 95° F)
Deviation in temperature can result in performance degradation.
Storage: 0° to 40° C (32° to 104° F) in original shipping packaging
Humidity
Operating: 20% to 60% RH
Storage: 10% to 70% RH, non-condensing in original packaging. The machine
should be packaged with desiccant if it is to be stored for an extended time.
Altitude
Operating: 0 m to 2,000 m (0 ft to 6,562 ft) above sea level
Contact Aerotech if your specific application involves use above 2,000 m or below
sea level.
Vibration
Use the system in a low vibration environment. Excessive floor or acoustical
vibration can affect system performance. Contact Aerotech for information
regarding your specific application.
Protection
Rating
Not suited for dusty or wet environments (IP30 equivalent ingress protection
rating).
Use
Indoor use only.
1.1.1. Accuracy and Temperature Effects
Aerotech products are designed for and built in a 20°C environment. Moderate temperature
changes will affect the accuracy while extreme temperature changes could cause damage to the
machine. At a minimum, the environmental temperature must be controlled to within 1°C per hour
to ensure the safety of the hexapod. The severity of temperature effects on all specifications
depends on many different environmental conditions, including how the hexapod is mounted.
Contact the factory for more details.