9
5.
Processing
5.1
Cleaning/Disinfecting
Note
Always follow disposal and hygiene instructions and
regulations!
Manual cleaning/disinfecting
¾
Wipe the shell of the product with a lint-free cloth
moistened with a commercial disinfectant for wipe
disinfection.
¾
If necessary, after the appropriate exposure time,
wipe off any residues of the cleaning/disinfecting
agent with a lint-free cloth moistened with clean
water.
¾
Dry the product with a clean, lint-free cloth.
¾
Clean the contacts in the charging bay with
isopropanol or ethyl alcohol and a cotton swab. Do
not use corrosive chemicals.
¾
If necessary, repeat the cleaning/disinfecting
process.
CAUTION
Damage to, or destruction of the
product caused by mechanical
cleaning/disinfecting or steriliza-
tion!
¾
Only clean and disinfect the
product manually.
¾
Do not sterilize the product
under any circumstances.
DANGER
Risk of electric shock and fire
hazard!
¾
Unplug the unit before cleaning.
¾
Make certain that no fluids will
penetrate the product.
¾
Leave the product exposed to
air for at least 1 minute after
cleaning/disinfecting.
CAUTION
Damage to the product due to
incorrect cleaning or inappropriate
cleaning/disinfecting agents!
¾
Use any commercial detergent
for surface cleaning.
¾
Never immerse the product in
water or detergent.
¾
Do not introduce any fluid into
the charging bays.
¾
Clean the contacts in the
charging bays, applying extreme
care.