46
turn on, or if the lancing device or adapters do not work after
cleaning and disinfection, discontinue use of the system and
contact Customer Service.
Availability of disinfecting wipe may vary. For updated informa-
tion, call Customer Service.
Table 1
EPA #
Brand Name
Manufacturer
Available at
Contact Time
9480-4*
Super
Sani-Cloth
Germicidal
Disposable Wipe
Professional
Disposables
International,
Inc. (PDI)
Online:
Amazon.com
Officedepot.com
2 Minutes
*Active ingredient: Quaternary ammonium chlorides and
isopropanol
How do you clean and disinfect your meter and lancing
device?
[1]
Before you disinfect your meter and lancing device, clean
the outside of the meter and lancing device with a disinfect-
ing wipe listed in the table above.
Wipe each side of your meter and lancing device with moder-
ate pressure 3 times using the following method:
a.
Wipe up and down 3 times
b.
Wipe left and right 3 times
c. Avoid getting disinfectant inside the test strip port.
[2]
With a new disinfecting wipe, disinfect your meter and
lancing device using the disinfecting wipe listed in the table