18
19
bottle. Discard the first drop of control
solution and wipe the bottle tip clean
using a clean tissue or paper towel.
Dispense a second drop onto a clean
surface, such as an unused resealable
plastic bag.
Bring the tip of the test strip to the control solution sample.
The AgaMatrix Jazz System will automatically detect if a test
is performed with AgaMatrix Control Solution.
When control solution is applied, the visual fill
window on the AgaMatrix Jazz Test Strip turns
blue. Your meter will beep once indicating that
you should remove the tip of the AgaMatrix Jazz
Test Strip from the control solution sample. On
the meter display, the droplet symbol fills. The
scrolling bar and 1-2-3 indicates that the meter is
calculating your results.
IMPORTANT: Count forward 90 days from the date you
open a new bottle of AgaMatrix Control Solution. This is
your discard date. Write the discard date on the AgaMatrix
Control Solution bottle.
Important AgaMatrix Control Solution information:
[1]
Use only AgaMatrix Control Solution with your AgaMatrix
Jazz System.
[2]
Always replace the cap on the control solution bottle
immediately after use.
[3]
Do not add water or any other liquid to control solution.
[4]
Control solution tests should be performed within the system
operating temperature range of 50 °F to 104 °F (10 °C to 40 °C).
How to perform a control solution test:
[1] Inserting an AgaMatrix
Jazz Test Strip:
Insert a new
AgaMatrix Jazz Test Strip into the
meter. The meter will be activated
when the test strip is fully inserted.
[2] Meter Ready for Test:
The meter now displays the
droplet symbol and the word “Fill”. The AgaMatrix Jazz
System is now ready for you to apply control solution. You do
not need to set a calibration code for this meter.
[3] Applying Control Solution:
Shake the control solution
This is the visual fi ll window.
DO NOT PUT CONTROL
SOLUTION HERE.
Apply control solution here.
The entire
visual fi ll
window will
turn blue.
Droplet
Symbol: