QC test image for mammography application
The QC image has several additional fields where the QC data can be filled in.
This image should be filed as part of the QC procedure.
The following two examples illustrate a QC test image for a different film
format.
Note
:
The dimensions A and B are not indicated on the printed
film. The QC objects on printed films may look different than in
the illustration. The location, size and shape of the QC objects
do not jeopardize the functionality of the QC and can be altered
depending on software versions, film types and film sizes.
• QC test image for film format 14x17”
Lo D
Base + F
og
M
ax D
Hi D
Mid D
Lo D
Sets of dots visible?
Yes
Base
+ Fog
DD
Density Difference
(
Hi D - Lo D)
Weekly spatial
Resolution test
Daily density tests
Geometry test
Max D
Hi D
Mid D
Identification
Quality Control Test Image
Initials:
Date:
Time:
No
2
3
1
• QC test image for film format 8x10”
Crystal-M Imager | Advanced Operation (key-operator mode) |
109
3852B EN 20210601 1655