3.
Record these values as measured dimensions A and B on Chart 4
(‘Geometric Consistency Control Chart’).
4.
Compare the measured A and B dimensions with the reference dimension
values, A
ref
and B
ref
on Chart 4 (‘Geometric Consistency Control Chart’).
The differences between measured dimensions of A and B and the
reference values A
ref
and B
ref
should be less than or equal to 1.0%.
5.
Check for image distortion.
6.
Calculate the aspect ratio by dividing A by B.
The result must be 1 +/- 0.01
WARNING:
If the image size or distortion values are outside of limits,
contact your local service organization to resolve the problem.
Related Links
Establishing the image geometry reference values
on page 100
Dry 2.4M | Advanced Operation (key-operator mode) |
105