JR6: Mechanical Setting
3. Remove sample holder and check if the axis of main shaft is in the
middle of the circle plate, as it is marked by the green circle in Fig.
.
4. If this not the case adjust it according to the next step. But do it
only, when the main shaft is far from the center of the hole!
Figure 9: Main shaft in correct position
5. Loose three screws marked by the red circles (only two of them
are marked in Fig.
) and move entire coil system to get the main
shaft in the center of the hole.
6. Do not forget to tighten the screws after successful adjustment.
Again - do this only, when the main shaft is far from the center of
the hole!
Service note
page 7