62
Agilaire 8872 Manual
Chapter 2 Configuration
Automatic Calibration configuration in Configuration Editors > Logger Channels
Configuring Automatic Calibrations
If you configure
Automatic
Calibrations, the cal you enable will be automatically initiated
by the data logger’s internal clock. The Automatic Cal Sequence configuration screen has the
following fields :
Calibration Type
will be already filled in (Automatic_A).
Calibration Name
is required to identify the cal program.
Check
Enabled
if the calibration is to run.
Number of Calibration Records
determines how many cals the data logger will store
before overwriting.
Recovery Time
specifies the time required to purge cal gas after phases.
Repeated Interval
determines how often cal sequence will repeat.
Start Time
determines what time cal sequence will start.
Affected Channels
determines which channels will be taken off-line during cal.
Select from a list of previously configured parameters.