Configuring User-Initiated Calibrations
User-initiated calibrations are started manually by linking to the data logger. When the cal
sequence is started, each phase will be initiated in order. The duration of each phase is
configured with the same fields as automatic cals.
The User-Initiated Cal configuration screen has the following fields :
Calibration Type
will be already filled in (UserInitiated_U).
Calibration Name
is required to identify the cal program.
Check
Enabled
if the calibration is to run.
Number of Calibration Records
determines how many cals the data logger will store
before overwriting.
Recovery Time
specifies the time required to purge cal gas after phases.
Affected Channels
determines which channels will be taken off-line during cal.
Select from a list of previously configured parameters.
Agilaire 8872 Manual
Chapter 2 Configuration
Configuring User-Initiated Calibrations