Characteristics
11
■
External magnetic field < 40 A/m
■
No adjacent current carrying conductor
■
1 M
Ω
/100 pF probe termination
■
Conductor centered in jaw
Table 2
Characteristics
Item
100 mV/A
Sensitivity Setting
10 mV/A
Sensitivity Setting
Current Measurement Range
0A to 10A dc or peak ac
0A to 100A dc or peak ac
Phase Shift (DC to 1 kHz,
no DC component)
< 1.5
°
dc to 65 Hz
< 1.0
°
dc to 65 Hz
Noise
3 mV
480 µV
Slew Rate
0.3 V/µs
20 mV/µs
Insertion Impedance (50 Hz/60 Hz)
0.01
Ω
0.01
Ω
Rise or Fall Time
3 µs
4 µs
Influence of Adjacent Conductor
<0.2 mA/A ac
Influence of Conductor Position in Jaw
0.5% of reading in jaw
Overload Indication
Red LED (OL) indicates input is out of range
Maximum Working Voltage (refer to
safety warnings and standards)
600Vac RMS CAT II
300Vac RMS CAT III
Maximum Floating Voltage (refer to
safety warnings and standards)
600Vac RMS CAT II
300Vac RMS CAT III
Table 3
Environmental and Mechanical Characteristics (Sheet 1 of 2)
Item
Characteristic
Environmental Conditions
Operating Temperature Range
0
°
C to 50
°
C
Storage Temperature Range
–30
°
C to 80
°
C
Temperature Influence
< 0.2% per
°
C
Max Operating Humidity
10
°
C to 30
°
C: 85% RH (without condensation)
40
°
C to 50
°
C: 45% RH (without condensation)
Altitude
Operating: 0 to 2000m
Non operating: 0 to 12,000m
Indoor Use
This probe is rated for indoor use only
Mechanical
Zero Adjustment
20 turn potentiometer
Maximum cable diameter
11.8 mm
Case Protection
IP20 per IEC 529
Drop Test
1.0 m on 38 mm of oak on concrete, tested according to IEC 61010
Mechanical Shock
100 G; test per IEC 68-2-27