background image

Performance Verification

15

Performance Verification

To ensure that the probe complies with the factory specifications, 
perform the six tests in this section in the order listed. Connect the 
required equipment listed in 

Table 5

 as shown in 

Figure 6 on page 

16

.

NOTE

Allow 30 minutes for warm up before starting the tests.  While the test system is 
warming up, clean the magnetic contacts on the probe’s jaw.

Construct a 10- turn coil with a diameter of about 4 inches (10 cm) 
as shown in 

Figure 5

. Use wire wrap or transformer wire and 

attach stackable banana plugs to the coil for connection to the 
instruments during testing.

Figure 5

10 Turn Coil

Table 5

Required Equipment for Performance Verification Tests

Description

Critical Specifications

Recommended 

 

Model/Part Numbers

Digital Multimeter

 

(2 required)

AC/DC voltage and current 
measurement accuracy better than 
±0.1% of reading Input resistance in 
AC/DC voltage mode 

¡

 1M

Ω

Agilent 34401A  or Agilent 3458A

Signal Generator

DC to 100 kHz sine waves

 

Able to generate more than 10 mA in 
the test coil (e.g. an inductive load).

Agilent 33120A or Fluke 5500A

Patch Cable Assembly

Banana plug connectors

 

Length 

¤

 36 in (91cm)

Pomona 1440-36-0

Adapter

BNC (m) to dual banana

Pomona 1296

Adapter

BNC (f) to dual banana

Pomona 1269

Banana Jack (2 required)

Stackable

Pomona 1325-0

Coil

10 turns  (requires about 4 ft or 1.2 m 
of transformer wire or wire wrap 
wire.)

OK Industries R30B-0100(100 ft roll of 
30AWG wire wrap wire)

Summary of Contents for 1146B

Page 1: ...Agilent 1146B AC DC Current Probe User s Guide...

Page 2: ...s or implied with regard to this manual and any information contained herein including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent shall n...

Page 3: ...afety Standards 12 Typical Response Curves 13 Performance Verification 15 If the Probe Fails a Test 16 Test 1 DC Current Measurement Accuracy 10 mV A range 17 Test 2 AC Current Measurement Accuracy 10...

Page 4: ...ital oscilloscope or other voltage measuring instrument which has the following features BNC input connector Range capable of displaying 0 2V to 0 5V per division Minimum input impedance of 1 M Before...

Page 5: ...uminates during measurement this indicates that the peak value exceeds the instrument response level and that the output is distorted Use the range selection switch to change the probe to a higher ran...

Page 6: ...an input channel on the oscilloscope WARNING Connect the probe to the oscilloscope or voltage measuring instrument before clamping the probe around a conductor 3 Set the current probe to its least sen...

Page 7: ...due WARNING Make sure that the probe is completely dry before reconnecting it to a power source Product Markings This symbol indicates the Environmental Protection Use Period EPUP for the product s to...

Page 8: ...be around a conductor WARNING Never use the probe on circuits rated higher than 600 Vac RMS CAT II or 300 Vac RMS CAT III or with float voltage greater than 600 V WARNING Never leave the probe clamped...

Page 9: ...on WARNING If you energize the instrument by an auto transformer for voltage reduction or mains isolation the ground pin of the input connector terminal must be connected to the earth terminal of the...

Page 10: ...stics are the typical performance values and are not warranted Characteristics are based on these conditions Within one year of calibration 23 C 5 C 20 to 75 relative humitidy Probe zeroed 1 minute wa...

Page 11: ...mum Working Voltage refer to safety warnings and standards 600Vac RMS CAT II 300Vac RMS CAT III Maximum Floating Voltage refer to safety warnings and standards 600Vac RMS CAT II 300Vac RMS CAT III Tab...

Page 12: ...68 2 6 frequency range 10 Hz to 55 Hz amplitude 0 15 mm Handle Lexan 920A UL 94 V2 Dimensions 231 mm x 36 mm x 67 mm Weight 330 g 11 6 oz with battery Color Light gray Output cable Insulated coaxial c...

Page 13: ...Characteristics 13 Typical Response Curves Figure 2 Typical Frequency Response Curve Figure 3 Typical Phase Shift Response Curve...

Page 14: ...Characteristics 14 Accuracy Figure 4 Linearity for a DC Signal Range 10 mV A...

Page 15: ...ure 5 10 Turn Coil Table 5 Required Equipment for Performance Verification Tests Description Critical Specifications Recommended Model Part Numbers Digital Multimeter 2 required AC DC voltage and curr...

Page 16: ...that The probe s magnetic contact surfaces are clean The probe tip closes completely around the coil s conductors The coil is centered in the probe tip opening The probe tip is perpendicular to the c...

Page 17: ...irection in the coil matches the current direction arrow on the probe Position the probe as perpendicular as possible to the coil 4 Set up the Agilent 33120A as follows Waveform Sine Frequency 1 kHz A...

Page 18: ...sary to achieve a current of at least 10 mA 11 Measure the voltage output by the probe VPROBE on DMM 2 and record in Table 7 12 Do the calculations specified in Table 7 to determine if the probe passe...

Page 19: ...be passes the test Table 7 AC Current Measurement Accuracy Record 10 mV A range Measurement Recorded Value Current in Coil IC Probe Output Voltage VPROBE Calculate Maximum Probe Output Voltage As VPRO...

Page 20: ...t direction in the coil matches the current direction arrow on the probe Position the probe as perpendicular as possible to the coil 20 Set up the Agilent 33120A as follows Waveform Sine Frequency 1 k...

Page 21: ...amplitude if necessary to achieve a current of at least 10 mA 27 Measure the voltage output VPROBE by the probe on DMM 2 and record in Table 10 28 Do the calculations specified in Table 10 to determin...

Page 22: ...he probe passes the test Table 10 AC Current Measurement Accuracy Record 100 mV A range Measurement Recorded Value Current in Coil IC Probe Output Voltage VPROBE Calculate Maximum Probe Output Voltage...

Page 23: ...I indicator lights 5 K knob 5 L Linearity for a DC signal Range 14 O OL 5 Operating 11 operating temperature range 11 P Peak Overload OL Indicator 5 peak ratings 5 probe cleaning 7 R range selection s...

Page 24: ...01146 92005 Agilent Technologies Manual Part Number 01146 92005 Printed in Malaysia...

Reviews: