24
Agilent 16441A User’s Guide, Edition 4
User's Guide
Specifications
Specifications
The "supplemental information" and "typical" entries, in the following
specifications are not warranted, but provide useful information about the functions
and performance of the instruments.
The following specifications data is specified at 23 ± 5
°
C and 50 % relative
humidity.
•
Function
Agilent 16441A R-BOX adds a selectable series resistor to the SMU output. You
can select the resistor from the setup page, and the voltage drop due to the series
resistor is automatically calculated. There are the limitations on measurement
with Agilent 4155/4156 Semiconductor Parameter Analyzers and R-Box:
•
If you measure device characteristics including negative resistance over
1 M
W
with the 4155/4156 and R-Box, there is a possibility that they cannot
measure it.
•
There is a possibility that the 4155/4156 cannot perform measurement
circumstances.
•
Resistance value and accuracy
Number of SMU channels that can add resistor: 2
Following resistance values are available for each channel (CH1, CH2):
•
1 M
W
•
100 k
W
•
10 k
W
•
0
W
resistance accuracy
0.3 % between Input and Output terminals
(at 23 ± 5
°
C and 50 % relative humidity)
•
Kelvin connection
Kelvin connection is effective only when 0
W
is selected.