Markers Field
The Markers field is used to access the markers selection pop-up. When the
Markers field is selected, a marker selection pop-up menu appears. The
choices are Sample (the default), Pattern, Time, and Statistics. Sample
markers are explained in more detail in the following section; pattern, time,
and statistics markers are explained in the "Markers and Stop Measurement"
chapter..
Markers Field
Sample Markers
The Sample selection turns X and O marker operations off, and enables the
X1 and X2 markers. The data values at the X1 and X2 markers are
automatically displayed to the right of the Marker field when charting label
versus state. Data between the X1 and X2 markers may be rescaled. For
more information on rescaling, refer to the "Rescale Field" later in this
chapter.
If a Compare Stop measurement is specified in the Stop measurement field,
and the Stop measurement criteria are met, the measurement will stop even
though the Markers are set to Sample. However, an X-O Stop measurement
will not complete if Markers are set to Sample.
Markers field
The Chart Menu
Markers Field
8–5
Summary of Contents for 16517A
Page 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Page 5: ...iv ...
Page 11: ...Contents 6 ...
Page 12: ...1 General Information ...
Page 18: ...2 Probing ...
Page 29: ...2 12 ...
Page 30: ...3 The Format Menu ...
Page 42: ...4 The Trigger Menu ...
Page 79: ...4 38 ...
Page 80: ...5 The Listing Menu ...
Page 86: ...6 The Waveform Menu ...
Page 106: ...7 The Compare Menu ...
Page 122: ...8 The Chart Menu ...
Page 133: ...8 12 ...
Page 134: ...9 Markers and Stop Measurements ...
Page 149: ...9 16 ...
Page 150: ...10 Skew Adjust ...
Page 156: ...11 Error Messages ...
Page 165: ...11 10 ...
Page 166: ...12 Specifications and Characteristics ...
Page 176: ...13 Installation and Service ...
Page 189: ...13 14 ...
Page 193: ...Index 4 ...