Skew Adjust
This chapter gives you instructions for calibrating and adjusting the
16517A/18A module. Though the specifications are not tested during
these procedures, the module must be adjusted to ensure the
specifications will be met.
The adjustment to the high-speed logic analyzer consists of skew
adjustment to minimize channel-to-channel skew. The adjustment
factors are stored in nonvolatile RAM, and are loaded up at system
start up.
To periodically verify the performance of the 16517A/18A against the
module specifications, refer to "Testing Performance" in the optional
Agilent Technologies 16517A/18A Service Guide
, available from
your Agilent Technologies Sales Office.
To Calibrate the Data Acquisition Module
The 16517A/18A does not require an operational accuracy calibration. To
test the module against the module specifications, refer to "Testing
Performance" in the optional
Agilent Technologies 16517A/18A Service
Guide
, available from your Agilent Technologies Sales Office.
10–2
Summary of Contents for 16517A
Page 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Page 5: ...iv ...
Page 11: ...Contents 6 ...
Page 12: ...1 General Information ...
Page 18: ...2 Probing ...
Page 29: ...2 12 ...
Page 30: ...3 The Format Menu ...
Page 42: ...4 The Trigger Menu ...
Page 79: ...4 38 ...
Page 80: ...5 The Listing Menu ...
Page 86: ...6 The Waveform Menu ...
Page 106: ...7 The Compare Menu ...
Page 122: ...8 The Chart Menu ...
Page 133: ...8 12 ...
Page 134: ...9 Markers and Stop Measurements ...
Page 149: ...9 16 ...
Page 150: ...10 Skew Adjust ...
Page 156: ...11 Error Messages ...
Page 165: ...11 10 ...
Page 166: ...12 Specifications and Characteristics ...
Page 176: ...13 Installation and Service ...
Page 189: ...13 14 ...
Page 193: ...Index 4 ...