Edge Width:
16/32/48/64/80 channels.
Edge Counting Frequency:
444 MHz.
Edge Detection:
Up to 1 GHz.
Greater than Duration (Timing only):
0 ns – 510 ns, accuracy is
±
2.25 ns.
Less than Duration (Timing only):
4 ns – 510 ns, accuracy is
±
2.25 ns.
Qualifier:
A user-specified term that can be any state, no state, any
recognizer (patterns or edges), the timer, or the logical combination
(AND, OR, XOR) of the recognizers and timers.
Branching:
Each sequence level has two branching qualifiers. When
satisfied, the analyzer will branch to the specified sequence level.
Maximum Occurrence Count:
16,777,216.
Maximum Sequencer Speed:
500 MHz.
State Sequence Levels:
4 plus trace point.
Timing Sequence Levels:
4 plus trace point.
Timer/Counter:
There is one timer or counter per sequence level,
which is restarted upon entry into each level.
Timer/Counter Range:
Timing mode
0 s to 33 ms.
State mode
500 MHz to 1 GHz, (user clock period)
×
(2 to the 23rd).
below 500 MHz, (user clock period)
×
(2 to the 24th).
Timer Resolution:
Timing mode
2 ns.
State mode
above 500 MHz, 2
×
user clock period.
below 500 MHz, user clock period.
Timer Accuracy:
0.005% of timer value.
Specifications and Characteristics
Characteristics
12–7
Summary of Contents for 16517A
Page 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Page 5: ...iv ...
Page 11: ...Contents 6 ...
Page 12: ...1 General Information ...
Page 18: ...2 Probing ...
Page 29: ...2 12 ...
Page 30: ...3 The Format Menu ...
Page 42: ...4 The Trigger Menu ...
Page 79: ...4 38 ...
Page 80: ...5 The Listing Menu ...
Page 86: ...6 The Waveform Menu ...
Page 106: ...7 The Compare Menu ...
Page 122: ...8 The Chart Menu ...
Page 133: ...8 12 ...
Page 134: ...9 Markers and Stop Measurements ...
Page 149: ...9 16 ...
Page 150: ...10 Skew Adjust ...
Page 156: ...11 Error Messages ...
Page 165: ...11 10 ...
Page 166: ...12 Specifications and Characteristics ...
Page 176: ...13 Installation and Service ...
Page 189: ...13 14 ...
Page 193: ...Index 4 ...