Perform the Skew Adjustments and Tests
Valid skew factors must be present in order to access these tests. The Skew
Tests require valid skew factors to ensure that the tests will pass.
1
Touch Skew Adjustments and Tests.
2
Perform the skew adjust procedure. Touch Adjust Clock and Data
Path Skew. A BNC coax cable (Agilent Technologies 10503A) and
Calibration Pod (16517-63201) are needed. Follow the instructions
on the display. At the end of the Skew adjust, save the skew factors
to NV-RAM. For more information, refer to "To Adjust the
Channel-to-channel Skew" in chapter 10.
Installation and Service
Perform the Skew Adjustments and Tests
13–11