Using Macros to Create a Trigger Specification
To configure a trigger specification using trigger macros, follow the
procedure below.
1
From the Trigger menu, enter the desired sequence level through the
Modify Trigger field, or by selecting a sequence level number.
See Also
"Editing Sequence Levels" and "Modify Trigger Field" found earlier in this
chapter for information on accessing levels.
2
From within the sequence level, select the Select New Macro field.
Select New Macro Field
3
Scroll and highlight the macro you want, then select the Select field.
4
Select the appropriate assignment fields and insert the desired
predefined resource terms, numeric values, and other parameter
fields required by the macro. Select the Done field.
See Also
"Resource Terms" found later in this chapter for information on using
predefined resource terms.
Select new macro field
Assignment field
The Trigger Menu
Using Macros to Create a Trigger Specification
4–11
Summary of Contents for 16517A
Page 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Page 5: ...iv ...
Page 11: ...Contents 6 ...
Page 12: ...1 General Information ...
Page 18: ...2 Probing ...
Page 29: ...2 12 ...
Page 30: ...3 The Format Menu ...
Page 42: ...4 The Trigger Menu ...
Page 79: ...4 38 ...
Page 80: ...5 The Listing Menu ...
Page 86: ...6 The Waveform Menu ...
Page 106: ...7 The Compare Menu ...
Page 122: ...8 The Chart Menu ...
Page 133: ...8 12 ...
Page 134: ...9 Markers and Stop Measurements ...
Page 149: ...9 16 ...
Page 150: ...10 Skew Adjust ...
Page 156: ...11 Error Messages ...
Page 165: ...11 10 ...
Page 166: ...12 Specifications and Characteristics ...
Page 176: ...13 Installation and Service ...
Page 189: ...13 14 ...
Page 193: ...Index 4 ...