If you cannot restore raw files
If the
setup.raw
and
data.raw
files seem to be ignored when you copy
them to the logic analyzer:
Verify the LAN connection between your computer and the logic
analyzer. Refer to "If you cannot connect to the logic analyzer" earlier in
this section.
Change the file time-out. To change the time-out:
1
Go to the logic analyzer System External I/O menu.
2
Select the LAN Settings field.
3
Set the file time out.
A value of about 1.5 seconds should be sufficient. Set the time out to
the minimum value that will allow network operations to work
reliably.
If you cannot telnet to the command parser socket
If you get a "connection refused" message:
Verify that you included the telnet port number (5025) in the command.
If you get a "connection timed out" or "no response from host" message:
Verify the LAN connection between your computer and the logic
analyzer. Refer to "If you cannot connect to the logic analyzer" earlier in
this section.
If you get a "connection refused" or "no response from host" message:
If the logic analyzer was just turned on, make sure that you wait 15
seconds after the Analyzer Configuration menu is displayed before
attempting the connection.
Make sure that another user is not already accessing the file system as
the control user or connected to the command parser socket.
Troubleshooting
If you cannot restore raw files
7–24
Summary of Contents for 1660 CS Series
Page 7: ...1 Connecting and Configuring...
Page 16: ...1 10...
Page 17: ...2 Accessing the Logic Analyzer File System...
Page 25: ...3 Using the X Window Interface...
Page 37: ...4 Retrieving and Restoring Data...
Page 50: ...4 14...
Page 51: ...5 Programming the Logic Analyzer...
Page 64: ...5 14...
Page 65: ...6 Concepts...
Page 72: ...6 8...
Page 73: ...7 Troubleshooting...
Page 104: ...Index Index 4...