If you get an "operation timed-out" message
Check the LAN connection between the computer andlogic analyzer.
Refer to "If you cannot connect to the logic analyzer" in this section.
Increase the file time-out value on your PC or workstation.
If the logic analyzer begins to operate slowly
The logic analyzer may operate more slowly if multiple users are trying to use
the system at the same time.
Check the number of network connections. Either copy the file
\system\mount.txt
to your local computer and view it with a text
editor, or on the analyzer go to the LAN Settings menu and select Show
LAN Connections. Check to see if many other users are currently
connected to the system.
If the logic analyzer does not respond
When the logic analyzer is transmitting a large file such as a deep-
memory acquisition or a screen image, or when it is transmitting a
dynamic file, it does not respond. Please be patient.
If the analyzer does not respond after several minutes, cycle the power.
For Agilent Technologies 1670-series logic analyzers with 512K memory
depth, data may take up to ten minutes to transmit. For most other
situations, three to five minutes is definitely long enough.
If all else fails
Contact your system administrator.
If you still cannot solve the problem, contact an Agilent Technologies
Service Center for repair information.
Troubleshooting
If you get an "operation timed-out" message
7–25
Summary of Contents for 1660 CS Series
Page 7: ...1 Connecting and Configuring...
Page 16: ...1 10...
Page 17: ...2 Accessing the Logic Analyzer File System...
Page 25: ...3 Using the X Window Interface...
Page 37: ...4 Retrieving and Restoring Data...
Page 50: ...4 14...
Page 51: ...5 Programming the Logic Analyzer...
Page 64: ...5 14...
Page 65: ...6 Concepts...
Page 72: ...6 8...
Page 73: ...7 Troubleshooting...
Page 104: ...Index Index 4...