Troubleshooting Flowchart 2
Is the RUN
indicator on?
Are the instrument
fans running?
Does the test
pass?
Is the cable
seated?
D o t h e L C D
display signals
test in Chapter 5.
E N D
2
R e m o v e p o w e r
f r o m t h e
i n s t r u m e n t .
C h e c k t h a t t h e
video data cable
i s s e a t e d o n C P U
b o a r d a n d o n
L C D d i s p l a y .
R e p l a c e C P U
B o a r d i n t h e
i n s t r u m e n t
R e p l a c e L C D
M o n i t o r
Is the power
suply fan
running?
E n s u r e t h e p o w e r
cord is properly
c o n n e c t e d .
Is the power cord
properly connected?
R e c o n n e c t t h e
p o w e r c o r d
S u s p e c t t h e p o w e r
s u p p l y . D o t h e p o w e r
s u p p l y v o l t a g e c h e c k
i n C h a p t e r 5 .
Do all test points
pass?
R e p l a c e p o w e r
s u p p l y
R e p l a c e
i n s t r u m e n t f a n
S e a t t h e v i d e o d a t a
c a b l e o n C P U b o a r d
a n d o n L C D d i s p l a y .
C H A R T 2
C H A R T 2
Y e s
Y e s
Y e s
No
Y e s
Y e s
Y e s
No
No
Y e s
No
No
No
No
Troubleshooting
To use the flowcharts
5–4
Summary of Contents for 1670G Series
Page 20: ...1 12...
Page 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Page 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Page 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Page 201: ...Theory of Operation The Oscilloscope Board 8 11...