Troubleshooting Flowchart 6a
E N D
6 a
P e r f o r m o s c i l l o s c o p e
o p e r a t i o n a l a c c u r a c y
calibration. Refer to
C h a p t e r 4 .
C H A R T 6
C H A R T 6 a
F i x e d ?
A t t e m p t t o d o a n
o s c i l l o s c o p e
acquisition.
R e p l a c e
o s c i l l o s c o p e b o a r d .
D o e s t i m e b a s e
s w e e p ?
C h e c k o s c i l l o s c o p e
p r o b e s .
R e m o v e p o w e r a n d
reseat oscilloscope
board interface
c a b l e .
Are oscilloscope
p r o b e s O K ?
Is data correct?
Does the
callibration pass?
R e p l a c e
o s c i l l o s c o p e p r o b e s .
R e p l a c e
o s c i l l o s c o p e b o a r d .
N o
Y e s
N o
N o
Y e s
Y e s
N o
N o
Y e s
Y e s
Troubleshooting
To use the flowcharts
5–9