Preparing For Use
This chapter gives you instructions for preparing the logic analyzer for use.
Power Requirements
The logic analyzer requires a power source of either 115 Vac or 230 Vac, –22 % to
+10 %, single phase, 48 to 66 Hz, CAT II pollution degree 2, 200 Watts maximum
power.
Operating Environment
The operating environment is listed in chapter 1. The logic analyzer will operate at
all specifications within the temperature and humidity range given in chapter 1.
However, reliability is enhanced when operating the logic analyzer within the
following ranges:
•
Temperature: +20
°
C to +35
°
C (+68
°
F to +95
°
F)
•
Humidity:
20% to 80% noncondensing
Note the recommended noncondensing humidity. Condensation within the
instrument can cause poor operation or malfunction. Provide protection against
internal condensation.
Storage
Store or ship the logic analyzer in environments within the following limits:
•
Temperature: -40
°
C to +75
°
C
•
Humidity:
Up to 90% at 65
°
C
•
Altitude:
Up to 15,300 meters (50,000 feet)
Protect the logic analyzer from temperature extremes which cause condensation on
the instrument.
To inspect the logic analyzer
1
Inspect the shipping container for damage.
If the shipping container or cushioning material is damaged, keep them until you have
checked the contents of the shipment and checked the instrument mechanically and
electrically.
2
Check the supplied accessories.
Accessories supplied with the logic analyzer are listed in "Accessories" in chapter 1.
3
Inspect the product for physical damage.
Check the logic analyzer and the supplied accessories for obvious physical or mechanical
defects. If you find any defects, contact your nearest Agilent Technologies Sales Office.
Arrangements for repair or replacement are made, at Agilent Technologies’ option, without
waiting for a claim settlement.
2–2
Summary of Contents for 1670G Series
Page 20: ...1 12...
Page 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Page 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Page 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Page 201: ...Theory of Operation The Oscilloscope Board 8 11...