c
Adjust the pulse generator channel 1 Delay until the pulses are aligned according the
the setup time of the setup/hold combination selected, +0.0 ps or -100 ps.
Disregard the oscilloscope Period(2) value. The settings provided in this procedure
measure the period from rising edge to rising edge, which is not a valid measurement.
10
Select the clock to be tested.
a
In the Master Clock menu, select the clock field to be tested, then select the clock
edge as indicated in the table. The first time through this test, use the top clock and
edge.
Clocks
J
↓
K
↓
L
↓
M
↓
b
Connect the clock to be tested to the pulse generator channel 1 output.
c
Select Done to exit the Master Clock menu.
11
Press the blue shift key, then press the Run key. If the analyzer obtains two to four
acquisitions without the "Stop Condition Satisfied" message appearing, then the test
passes. Press Stop to halt the acquisition. Record the Pass or Fail results in the
performance test record.
Testing Performance
To test the single-clock, single-edge, state acquisition (logic analyzer)
3–27