Acquire the data
Use the following table for steps 1 through 5.
Oscilloscope Settings
Voltage Readings
V/Div
Offset
Supply
Upper Limit
Lower Limit
4 V/Div
-7.0 V
-14.0 V
-13.7 V
-14.3 V
1 V/Div
-1.75 V
-3.50 V
-3.43 V
-3.57 V
400 mV/Div
-700.0 mV
-1.40 V
-1.37 V
-1.43 V
40 mV/Div
-70.0 mV
-140.0 mV
-137.0 mV
-143.0 mV
40 mV/Div
70.0 mV
140.0 mV
143.0 mV
137.0 mV
400 mV/Div
700.0 mV
1.40 V
1.43 V
1.37 V
1 V/Div
1.75 V
3.50 V
3.57 V
3.43 V
4 V/Div
7.0 V
14.0 V
14.3 V
13.7 V
1
Set up the oscilloscope according to the table above.
a
Press the Chan key.
b
Move the cursor to the V/Div field, then use the RPG knob to dial in the V/Div values
shown on the first line of the table.
c
Move the cursor to the Offset field. Use the front-panel keyboard to enter the offset
value shown on the first line of the table. Use the left and right cursor-control keys to
select either mV or V.
2
Acquire the zero input voltage.
a
Disconnect the power supply from the channel input.
b
Press the blue shift key, then press the Run key. Wait for approximately five seconds
(averaging complete), then press Stop.
c
Press the Markers key. Move the cursor to the Va Volts field. Using the RPG knob,
move the Va marker over the oscilloscope trace on the display.
3
Acquire the measured voltage.
a
Connect the power supply to the channel input. Set the power supply voltage
according to the first line of the table above.
b
Press the blue shift key, then press the Run key. Wait for approximately five seconds
(averaging complete), then press Stop.
c
Press the Markers key. Move the cursor to the Vb Volts field. Using the RPG knob,
move the Vb marker over the oscilloscope trace on the display.
d
Record the voltage in the Va to Vb field in the performance test record.
4
Repeat steps 1 through 3 for the second line of the table, then for the rest of the lines
of the table for channel 1.
5
Repeat from "Set up the logic analyzer" for channel 2, replacing channel 1 with
channel 2 where applicable.
Testing Performance
To test the voltage measurement accuracy (oscilloscope)
3–66
Summary of Contents for 1670G Series
Page 20: ...1 12...
Page 116: ...Testing Performance Performance Test Record pattern generator 3 92...
Page 126: ...Calibrating and Adjusting To test the CAL OUTPUT ports 4 10...
Page 166: ...Exploded View of the Agilent 1670G series logic analyzer Replacing Assemblies 6 4...
Page 201: ...Theory of Operation The Oscilloscope Board 8 11...