3
Set up the Trigger menu.
a
Press the Trigger key.
b
Select the Mode/Arm field, then select Immediate.
4
Set up the Marker menu.
a
Press the Marker key.
b
Move the cursor to the T Markers field. Press Select, and then press On.
c
If the V markers are On, turn the V markers Off by moving the cursor to the V markers
field and pressing Select.
Connect the logic analyzer
1
Using a BNC-to-banana adapter, connect one end of the cable to the power supply.
Connect the BNC tee, the blocking capacitor, and the shorting endcap to the other
end of the cable.
2
Monitor the power supply output with the Digital Multimeter.
Testing Performance
To test the offset accuracy (oscilloscope)
3–69