107
Chapter 2: Task Guide
Cross-Triggering
2. Run the measurement.
To cross-trigger with another instrument
1. Select the Intermodule button (or from the Window menu, select System
and Intermodule).
2. In the Intermodule window, select the icon of the instrument to be armed,
and choose the instrument that will arm it.
When the logic analyzer waits for the arm signal
1. In the Trigger tab's Trigger Sequence area, select the sequence level that
should wait for the other analyzer's trigger.
2. In the
Trigger Functions
subtab, select the
Wait for arm in
trigger
function and select either the
Replace
or
Insert before
button.
(You can also use an advanced trigger function, edit it, and insert an
Arm
in from IMB
event.)
NOTE:
If the trigger sequence does not pass through the level containing the
wait for
arm
event, the logic analyzer will not wait for the arming signal.
When the logic analyzer drives the arm signal
1. In the Trigger tab's Settings subtab, select the analyzer (or analyzers) that
will be used to drive the arm signal.
2. Set up the logic analyzer trigger as you would normally.
In the Trigger tab's Trigger Sequence area, the trigger actions will show
arm out
to indicate that the logic analyzer's trigger is driving the arm
signal.
3. Run the measurement.
See Also
Intermodule Window (see the
Agilent Technologies 16700A/B-Series
Logic Analysis System
help volume)
Group Run Arming Tree (see the
Agilent Technologies 16700A/B-Series
Logic Analysis System
help volume)