141
Chapter 3: Reference
Importing Netlist and ASCII Files
than 5 nsec and the clock period is greater than 10 nsec (slower than
100 MHz).
Eye Finder Load/Save Messages.
These messages can appear when
saving or loading
eye finder
data.
"... (at line XX in the file)"
Indicates where the error occurred in the file being read. Since
eye
finder
data files are ASCII text, you can use a text editor to examine
the file at the indicated line to determine how to repair the problem.
"Bad assignment for XXX"
The numerical value for the item XXX could not be read.
"Cannot load Eye Finder data for a different module type. Data in file is from a
OTHERMODEL. This module is a THISMODEL."
"Cannot load Eye Finder data for a module installed in different slots. Data in file is
from a module installed in slots B A C. This module is installed in slots E D F".
"Cannot load Eye Finder data from a different instrument. Data in file is from
OTHERINSTRUMENT. This instrument is THISINSTRUMENT."
"Channel name in the data file ("OTHERNAME") does not match the expected
channel name ("THISNAME")."
"Channel number in the data file (NN) does not match the expected channel number
(MM).",
"Did not find: XXX (Search began at line LLL)"
The
eye finder
data file has a nested block structure. The next block
(or item) to be read was XXX, but it was not found before the end of
the file. The line after the last item successfully read was LLL.
"Error in writing"
Disk is probably full.