189
Chapter 3: Reference
Specifications and Characteristics
40 degrees C (+104 degrees F)
- Altitude Operating, to 4600 m (15,000 ft)
Non-operating, to 15,300 m (50,000 ft)
- Vibration
Operating: Random vibration 5-500 Hz, 10 minutes per axis,
approximately 0.2 g rms
Nonoperating: Random vibration 5 to 500 Hz, 10 minutes per axis,
approximately 2.41 g rms; and swept sine resonant
search, 5 to 500 Hz, 0.50 g (0-peak), 5-minute
resonant dwell at 4 resonances per axis.
Reliability is enhanced when operating within the following ranges:
- Temperature +20 to 35 degrees C (+68 to 95 degrees F)
- Humidity 20% to 80% non-condensing
Storage
Store or ship the logic analyzer in environments with the following
limits:
- Temperature -40 to +75 degrees C
- Humidity up to 90% at 65 degrees C
- Altitude up to 15,300 meters (50,000 feet)
Protect the module from temperature extremes which cause condensation
on the instrument.
What is a Specification?
A
Specification
is a numeric value, or range of values, that bounds the
performance of a product parameter. The product warranty covers the
performance of parameters described by specifications. Products
shipped from the factory meet all specifications. Additionally, the
products sent to Agilent Technologies Customer Service Centers for
calibration and returned to the customer meet all specifications.
Specifications are verified by
Calibration Procedures
.
What is a Calibration
Procedure?
Calibration procedures verify that products or systems operate within
the specifications. Parameters covered by specifications have a
corresponding calibration procedure. Calibration procedures include
both performance tests and system verification procedure. Calibration
procedures are traceable and must specify adequate calibration
standards.
Calibration procedures verify products meet the specifications by
comparing measured parameters against a pass-fail limit. The pass-fail
limit is the specification less any required guardband.
The term "calibration" refers to the process of measuring parameters
and referencing the measurement to a calibration standard rather than
the process of adjusting products for optimal performance, which is
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com