68
Chapter 2: Task Guide
Setting Up Triggers and Running Measurements
"Branches taken" feature of past logic analyzers. The best way to store
only the states that cause sequence level branches is by setting up
default storing to
Nothing
, inserting a
Store sample
action in each
sequence level, and inserting a
Turn off default storing
action in the
level that triggers.
NOTE:
When store qualification is performed in the 333 MHz State mode, there may
be the case where data occupying memory is further disqualified. As a result,
you may see a non-contiguous listing of states as well as a reduction of usable
memory.
To clear default storing changes
1. When the Trigger tab is displayed, select
Clear Default Store
from the
Clear menu.
See Also
“Storage Qualification” on page 201 in “Understanding Logic Analyzer
Triggering” on page 192
“To insert a store action (state mode)” on page 79
“To specify whether default storing is initially on or off” on page 68
To specify whether default storing is initially on or off
In the
state
sampling mode, you can specify whether the default
storing is initially on or off.
1. In the Trigger tab's Default Storing subtab, select the
At start of
acquisition
option button and choose either
On
or
Off
.
See Also
“Storage Qualification” on page 201 in “Understanding Logic Analyzer
Triggering” on page 192
“To insert a store action (state mode)” on page 79
“To Specify Default Storing” on page 67
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com