Chapter 5: Troubleshooting
General Troubleshooting
e
Click the Master button for the clock to be tested; then, select Both
Edges.
5
Configure the pod under test:
a
In the Analyzer Setup dialog, click the Buses/Signals tab
b
Click Delete All at the bottom of the window.
c
Using the mouse, activate all channels for the pod under test. Assign
channels to bus/signal name My Bus 1.
d
Click on the threshold field for the pod under test. The Threshold
Settings dialog appears.
e
In the Threshold Settings dialog, select Standard and ECL (-1.30 V).
f
Click OK to close the the Threshold Settings dialog.
g
Click OK to close the Analyzer Setup dialog.
6
Verify the data:
a
Click the Listing - Listing 1 tab.
b
In the Listing window, click the
Run icon. The display should look