169
Chapter 8: Theory of Operation
Self-Tests Descriptions
Trigger Arm Test.
The Trigger Arm Test verifies that the local arm signal can be
received by the master acquisition IC on the acquisition board. The test also
verifies the global arm signal can be driven by each acquisition IC on a master
board and received by all acquisition ICs on the card. The arm lines are asserted
and read at the acquisition ICs to ensure each acquisition IC recognizes the
signal.
Passing the Trigger Arm Test implies any acquisition IC can arm the card and that
all acquisition ICs can recognize the arm signal.
Clock Paths Test.
The Clock Paths Test verifies that the system Master, Slave,
and Psync clocks are functional between the acquisition ICs. The module is
configured to take a simple measurement. Test data is then created at the
comparators and an acquisition taken. The resulting data is then downloaded and
compared with known values.
Passing the Clock Paths Test implies that all acquisition IC clock lines can be
driven by each acquisition IC and can be received by each acquisition IC in the
module. Consequently each acquisition IC can reliably acquire data in response to
the acquisition clock signal.
Memory Modes Test.
The Memory Modes Test verifies the CPU interface can
properly manage the acquisition memory unload in full channel, half-channel,
count only, and interleaved modes. Test data is written to acquisition memory.
Different unload modes are selected, then the data is read and compared with
known values.
Passing the Memory Modes Test implies that the data can be reliably read from
acquisition memory in full channel, half-channel, count only, and interleaved
mode. This test along with the Memory Test provides complete testing of
acquisition memory downloading through the 1394 interface.
Calibration Test.
The Calibration Test ensures that each acquisition IC in the
module can perform an operational accuracy self-calibration. Various self-
calibration routines are initiated. The results of each self-calibration routine are
then checked to see if the self-calibration was successful or not.
Passing the Calibration Test implies that the module can reliably perform an
operation accuracy self-calibration. Consequently the incoming data path is
optimized to reduce channel-to-channel skew so the acquisition ICs can reliably
capture the incoming data.
Logic Analyzer Self-Tests
Register Test.
The Register Test verifies that the registers of each acquisition IC
is operating properly. Test patterns are written to each register on each
acquisition IC, read, and compared with know values. The registers are reset, and
Summary of Contents for 1680 series
Page 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Page 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Page 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Page 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Page 174: ......