Chapter 3: Testing Performance
To set up the test equipment and the logic analyzer
b
Configure the oscilloscope according to the following table.
Oscilloscope Setup
Set up the 1680A,AD-series logic analyzer
Power-up self tests are done on the logic analyzer system components when
power is applied. Any problems reported by the logic analyzer during boot must
be cleared before going further. For more information, refer to Chapter 5 and
Chapter 8.
1
Turn on the logic analyzer:
a
Connect a keyboard and mouse to the rear panel of the logic analyzer.
b
Plug in a power cord to the rear panel power connector of the logic
analyzer.
c
Turn on the power switch on the logic analyzer front panel.
2
Set up the logic analyzer:
a
Wait for the logic analyzer boot up to complete.
b
On the logic analyzer desktop, double-click the Agilent Logic Analyzer
icon to launch the application.
Acquisition
Display
Trigger
[Shift]
∆
Time
Averaging: On
# of averages: 16
Graticule graphs: 2
Level: 0.0 V
Stop src: channel 2 [Enter]
Channel 1
Channel 2
Define meas
External Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset: - 1.300 V
External Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset: - 1.300 V
Thresholds: user-defined
Units: Volts
Upper: - 980 mV
Middle: -1.30 V
Lower: -1.62 V
Summary of Contents for 1680 series
Page 13: ...13 Chapter 1 General Information Dimensions 1680A AD Series 1690A AD Series...
Page 74: ...74 Chapter 3 Testing Performance Performance Test Record...
Page 142: ...142 Chapter 6 Replacing Assemblies 1690A AD series disassembly assembly...
Page 172: ...172 Chapter 8 Theory of Operation Self Tests Descriptions...
Page 174: ......